1. Product Introduction and Structure
PET Sensitivity Phantom
Length: 70 cm; Wall thickness: 1 mm; Comes standard with 2 line source tubes.
Scatter Fraction Phantom
Length: 70 cm; Diameter: 20 cm; Comes standard with 2 line source tubes.
PET Spatial Resolution Bracket
It can support the sensitivity phantom and capillary tubes for measuring the corresponding PET performance indicators. The coordinates of capillary insertion holes are (0,1 cm), (0,10 cm), (10 cm,0); the coordinates of sensitivity phantom insertion holes are (0,0), (0,10 cm). It features an adjustable horizontal design.
Performance Requirements for Portable Radionuclide Automatic Dispenser
1)Mass < 10 kg;
2)Fully electric control (non-pneumatic);
3)Dispensing realized by peristaltic pump;
4)Three-channel automatic control of radionuclides to be dispensed;
5)Fully automatic control of the dispensing process;
6)The activity meter reading can be displayed in real time in the control software.
Nuclear Medicine Image Quality Control Analysis Software
It analyzes DICOM format images and can calculate the intrinsic spatial uniformity, intrinsic spatial linearity, spatial resolution, system sensitivity, and PET/CT image fusion accuracy.
It can measure the following PET indicators: spatial resolution, sensitivity, scatter fraction, NECR; and provide technical training services for the quality control detection of nuclear medicine equipment. Accuracy: 1; Measurement range: 1-11.
2. Detection Performance
PET Phantom Multifunctional Bracket
It can support the sensitivity phantom and capillary tubes for measuring the corresponding PET performance indicators; it has an adjustable horizontal design to meet the precise positioning requirements of CT laser lights in X, Y and Z directions.
It meets the quality control requirements of PET standards WS817-2023 and NEMA-2018 for tests at 1 cm, 10 cm and 20 cm off-center.
It complies with the quality control requirements for (1,0), (10,0) and (20,0) specified in PET standards WS817-2023 and NEMA-2012.
It can be horizontally adjusted in forward and reverse directions, suitable for the detection conditions of various models, and ensures that the test phantom is parallel to the PET field of view center.
A.1. PET Sensitivity Phantom
A.1. Scattering Fraction Phantom
A.3. Point Source Holder for Spatial Resolution Detection

http://www.hkmedqc.com/uploadfile/2026/0401/20260401033625822.pdf